COVID-19: A flow-inflating device for providing continuous positive airway pressure for children in resource-limited setups


1 Department of Anesthesiology, Indira Gandhi Institute of Medical Sciences, Patna, Bihar, India
2 Department of Maxillofacial Surgery, All India Institute of Medical Sciences, Patna, Bihar, India
3 Department of Undergraduate, MGM Medical College, Navi Mumbai, Maharashtra, India

Date of Submission21-Feb-2022
Date of Decision09-Mar-2022
Date of Acceptance18-Mar-2022
Date of Web Publication28-Jul-2022

Correspondence Address:
Prof. Prakash K Dubey
E ¾, IGIMS Campus, Patna - 800 014, Bihar
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijrc.ijrc_38_22

Rights and Permissions

How to cite this article:
Dubey PK, Dubey P, Dubey N. COVID-19: A flow-inflating device for providing continuous positive airway pressure for children in resource-limited setups. Indian J Respir Care 2022;11:281-2

How to cite this URL:
Dubey PK, Dubey P, Dubey N. COVID-19: A flow-inflating device for providing continuous positive airway pressure for children in resource-limited setups. Indian J Respir Care [serial online] 2022 [cited 2022 Aug 11];11:281-2. Available from: http://www.ijrc.in/text.asp?2022/11/3/281/352645



Sir,

Continuous positive airway pressure (CPAP) may prove useful in pediatric COVID-19 patients by providing non-invasive ventilation (NIV). Positive pressure NIV, bubble CPAP, and bilevel positive airway pressure ventilation have been recommended to manage children infected with COVID-19.[1] Environmental safety concerns and effects on work of breathing when attaching filters in mechanical model with CPAP for use in COVID-19 pediatric patients have also been evaluated.[2]

Jackson-Rees circuit (Mapleson F system) has been used to alleviate respiratory failure in a wide variety of primary pulmonary disorders.[3],[4] Jackson-Rees circuit consists of an adjustable expiratory valve at the end of the reservoir bag. Partial closure of the valve during spontaneous respiration provides CPAP, while intermittent compression of the bag delivers positive pressure ventilation. We describe a modification in the Jackson-Rees circuit using an NIV mask that can be used for this purpose in situations where no other equipment is available.

The reservoir bag of the Jackson-Rees circuit is accommodated inside an appropriately sized transparent plastic container [Figure 1]. A tubing of breathing circuit is attached from the other end of the container to serve as an exhaust for the accumulated gases in the container. Now, all the exhaled gas emitting from the valve of the reservoir bag is collected within the container and is carried away through the exhaust tubing. The distal end can be dipped into an underwater seal system containing 1% sodium hypochlorite solution to inactivate the contaminated gas.[5] Alternatively, a bacterial and viral filter can be attached to the distal end to prevent environmental contamination.[2]
Figure 1: The assembled device with the yellow arrow shows the direction of oxygen flow, and the red arrow shows the direction of scavenging the exhaled gases

Click here to view


An NIV mask is attached to the child's face to cover the nose and mouth, providing an air-tight seal and fixed in position using the headgear. The inspiratory limb of the circuit is attached to an appropriate oxygen source. The valve of the reservoir bag is kept partially opened. Oxygen flow (three times the patient's minute volume) is started to prevent rebreathing. After observing oxygen saturation by a pulse oximeter, the valve is gradually closed to provide CPAP. It has been suggested that if the expiratory valve is kept partially open so that the reservoir bag is neither overinflated nor collapsed, it will produce CPAP.[3] The aim is to achieve an oxygen saturation of more than 95%. Now, the reservoir bag is disconnected and kept inside the container. The container's lid is screwed tightly after attaching the bag to the expiratory limb of the Jackson-Rees circuit so that all the exhaled gas is scavenged after being collected in the container.

Disconnection, airway obstruction, and apnea can be detected by observing the intermittent inflation and deflation of the reservoir bag. In the case of apnea, the container can be removed, and positive pressure ventilation can be instituted through the bag. A nasogastric tube can be placed if the child develops abdominal distension due to aerophagy.

The limitations of this system include the inability to titrate inspired oxygen concentration and patients in whom it is difficult to maintain the airway. A bystander clinician should always be present as quantification of oxygen delivered is impossible, and no backup respiratory rate can be set. CPAP is not a substitute for invasive ventilation, and prompt intubation must be performed when the need arises.

Keeping in view its utility and limitations, we suggest its use for stabilizing and transporting sick children during the pandemic surge.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Kache S, Chisti MJ, Gumbo F, Mupere E, Zhi X, Nallasamy K, et al. COVID-19 PICU guidelines: For high- and limited-resource settings. Pediatr Res 2020;88:705-16.  Back to cited text no. 1
    
2.
Donaldsson S, Naver L, Jonsson B, Drevhammar T. COVID-19: Minimising contaminated aerosol spreading during CPAP treatment. Arch Dis Child Fetal Neonatal Ed 2020;105:669-71.  Back to cited text no. 2
    
3.
Anitha GF, Velmurugan L, Sangareddi S, Nedunchelian K, Selvaraj V. Effectiveness of flow inflating device in providing Continuous Positive Airway Pressure for critically ill children in limited-resource settings: A prospective observational study. Indian J Crit Care Med 2016;20:441-7.  Back to cited text no. 3
[PUBMED]  [Full text]  
4.
Santhanam I. Flow inflating ventilation device: Non-invasive CPAP in settings without immediate access to mechanical ventilation. In: Pediatric Emergency Medicine Course. 2nd ed. New Delhi: Jaypee; 2013. p. 53-7.  Back to cited text no. 4
    
5.
Kumar A, Kumar N, Kumar A, Kumar A, Sinha C. Exhaled gas decontamination by connecting a ventilator exhaust port with an underwater seal system to prevent COVID-19 spread. Trends Anaesth Crit Care 2021;36:45-6.  Back to cited text no. 5
    


    Figures

  [Figure 1]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
References
Article Figures

 Article Access Statistics
    Viewed80    
    Printed0    
    Emailed0    
    PDF Downloaded6    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]